MODEL |
Na+(C-122) |
|
Principle |
Sodium ion electrode
method |
|
Display |
LCD, 2-digit |
|
Guarantee range |
23-2300 ppm
(10-3-10-1 mol/l)* |
|
Display range |
0-99 x 100 ppm |
|
Resolution |
1 ppm for 0-99 ppm
10 ppm for 10-99 x 10 ppm
100 ppm for 10-99 x 100 ppm |
|
Calibration |
2-point calibration by STD/SLOPE
controls using standard calibration solutions
STD:2000 ppm, SLOPE: 150 ppm |
|
Repeatability |
±20% of display value |
|
Sample temperature |
5-35°C/41-95°F
(temperature compensated at ambient temperature) |
|
Power |
CR-2025 lithium cell ( x 2)
approx.500 hrs continuous operation |
|
Weight |
Approx.40g, 1.4oz |
|
Dimensions |
95 x 55 x 9mm,3.7 x 2.2 x 0.35in |
|
Standard accessories |
Na+ standard
solution
STD 2000 ppm, SLOPE 150 ppm |
Deionized water, sampling sheet,
tweezers, pipet, instruction booklet |
* Units can be used beyond 10-3 to 10-1
mol/l.
| MODEL
|
K+(C-131) |
| Principle
|
Potassium ion electrode method |
| Display |
LCD, 2-digit |
| Guarantee
range |
9-3900 ppm
(10-3-10-1 mol/l)* |
| Display range
|
0-99 x 100 ppm |
| Resolution
|
1 ppm for 0-99 ppm
10 ppm for 10-99 x 10 ppm
100 ppm for 10-99 x 100 ppm |
| Calibration
|
2-point calibration by STD/SLOPE
controls using standard calibration solutions
STD:2000 ppm, SLOPE: 150 ppm |
| Repeatability
|
±20% of display value |
| Sample temperature
|
5-35°C/41-95°F
(temperature compensated at ambient temperature) |
| Power |
CR-2025 lithium cell ( x 2)
approx.500 hrs continuous operation |
| Weight |
Approx.40g, 1.4oz |
| Dimensions
|
95 x 55 x 9mm,3.7 x 2.2 x 0.35in |
|
Standard accessories |
K+ standard solution
STD 2000 ppm, SLOPE 150 ppm |
Deionized water, sampling sheet,
tweezers, pipet, instruction booklet |
* Units can be used beyond 10-3 to 10-1
mol/l.
| MODEL
|
NO3-(C-141) |
| Principle
|
Nitrate ion electrode method |
| Display |
LCD, 2-digit |
| Guarantee
range |
62-6200 ppm
(10-3-10-1 mol/l)* |
| Display range
|
0-99 x 100 ppm |
| Resolution
|
1 ppm for 0-99 ppm
10 ppm for 10-99 x 10 ppm
100 ppm for 10-99 x 100 ppm |
| Calibration
|
2-point calibration by STD/SLOPE
controls using standard calibration solutions
STD:2000 ppm, SLOPE: 150 ppm |
| Repeatability
|
±20% of display value |
| Sample temperature
|
5-35°C/41-95°F
(temperature compensated at ambient temperature) |
| Power |
CR-2025 lithium cell ( x 2)
approx.500 hrs continuous operation |
| Weight |
Approx.40g, 1.4oz |
| Dimensions
|
95 x 55 x 9mm,3.7 x 2.2 x 0.35in |
|
Standard accessories |
NO3- standard solution
STD 2000 ppm, SLOPE 150 ppm |
Deionized water, sampling sheet,
tweezers, pipet, instruction booklet |
* Units can be used beyond 10-3 to 10-1
mol/l.

CARDY Ion Meters'
Electrode Selectivity Coefficients
| |
Na+ |
K+ |
NO3- |
Selectivity
coefficient |
K+=1
x 10-2
Rb+=1 x 10-2
Li+=1 x 10-3
NH4+=6 x 10-3
Ca2+=1 x 10-4
Sr2+=1 x 10-4
Ba2+=1 x 10-4
Mg2+=1 x 10-4
Cs+=3 x 10-3
pH range:pH3-9
(at 10-3 mol/l)
|
Rb+=1
x 10-1
Na+=3 x 10-4
NH4+=7 x 10-3
Mg2+=1 x 10-5
Cs+=4 x 10-3
Ca2+=7 x 10-7
pH range:pH 2-9
(at 10-3 mol/l)
|
l-=7
x 10-10
Cl-=4 x 10-2
Br-=9 x 10-1
ClO4-=3 x 10-3
NO2-=7 x 10-1
pH range:pH 3-8
(at 10-3 mol/l)
|
|