Japan Worldwide
News & EventsInvestorsProcurementCompany HomeContactsSite MapJapanese
Products & Support Automotive Process & Environmental Medical-Diagnostics Semiconductor Scientific Applications
Measurement Types
Product List
 HOME >> Products & Support
[genre_e] Products
 
Compact Salt Meter
C-121

product image
C-121
Specifications
Details
Detailed Specifications
MODEL Na+(C-122)
Principle Sodium ion electrode method
Display LCD, 2-digit
Guarantee range 23-2300 ppm
(10-3-10-1 mol/l)*
Display range 0-99 x 100 ppm
Resolution 1 ppm for 0-99 ppm
10 ppm for 10-99 x 10 ppm
100 ppm for 10-99 x 100 ppm
Calibration 2-point calibration by STD/SLOPE
controls using standard calibration solutions
STD:2000 ppm, SLOPE: 150 ppm
Repeatability ±20% of display value
Sample temperature 5-35°C/41-95°F
(temperature compensated at ambient temperature)
Power CR-2025 lithium cell ( x 2)
approx.500 hrs continuous operation
Weight Approx.40g, 1.4oz
Dimensions 95 x 55 x 9mm,3.7 x 2.2 x 0.35in
Standard accessories Na+ standard solution
STD 2000 ppm, SLOPE 150 ppm
Deionized water, sampling sheet,
tweezers, pipet, instruction booklet
* Units can be used beyond 10-3 to 10-1 mol/l.

MODEL K+(C-131)
Principle Potassium ion electrode method
Display LCD, 2-digit
Guarantee range 9-3900 ppm
(10-3-10-1 mol/l)*
Display range 0-99 x 100 ppm
Resolution 1 ppm for 0-99 ppm
10 ppm for 10-99 x 10 ppm
100 ppm for 10-99 x 100 ppm
Calibration 2-point calibration by STD/SLOPE
controls using standard calibration solutions
STD:2000 ppm, SLOPE: 150 ppm
Repeatability ±20% of display value
Sample temperature 5-35°C/41-95°F
(temperature compensated at ambient temperature)
Power CR-2025 lithium cell ( x 2)
approx.500 hrs continuous operation
Weight Approx.40g, 1.4oz
Dimensions 95 x 55 x 9mm,3.7 x 2.2 x 0.35in
Standard accessories K+ standard solution
STD 2000 ppm, SLOPE 150 ppm
Deionized water, sampling sheet,
tweezers, pipet, instruction booklet
* Units can be used beyond 10-3 to 10-1 mol/l.

MODEL NO3-(C-141)
Principle Nitrate ion electrode method
Display LCD, 2-digit
Guarantee range 62-6200 ppm
(10-3-10-1 mol/l)*
Display range 0-99 x 100 ppm
Resolution 1 ppm for 0-99 ppm
10 ppm for 10-99 x 10 ppm
100 ppm for 10-99 x 100 ppm
Calibration 2-point calibration by STD/SLOPE
controls using standard calibration solutions
STD:2000 ppm, SLOPE: 150 ppm
Repeatability ±20% of display value
Sample temperature 5-35°C/41-95°F
(temperature compensated at ambient temperature)
Power CR-2025 lithium cell ( x 2)
approx.500 hrs continuous operation
Weight Approx.40g, 1.4oz
Dimensions 95 x 55 x 9mm,3.7 x 2.2 x 0.35in
Standard accessories NO3- standard solution
STD 2000 ppm, SLOPE 150 ppm
Deionized water, sampling sheet,
tweezers, pipet, instruction booklet
* Units can be used beyond 10-3 to 10-1 mol/l.


CARDY Ion Meters'
Electrode Selectivity Coefficients

  Na+ K+ NO3-
Selectivity
coefficient
K+=1 x 10-2
Rb+=1 x 10-2
Li+=1 x 10-3
NH4+=6 x 10-3
Ca2+=1 x 10-4
Sr2+=1 x 10-4
Ba2+=1 x 10-4
Mg2+=1 x 10-4
Cs+=3 x 10-3

pH range:pH3-9
(at 10-3 mol/l)

Rb+=1 x 10-1
Na+=3 x 10-4
NH4+=7 x 10-3
Mg2+=1 x 10-5
Cs+=4 x 10-3
Ca2+=7 x 10-7

pH range:pH 2-9
(at 10-3 mol/l)

l-=7 x 10-10
Cl-=4 x 10-2
Br-=9 x 10-1
ClO4-=3 x 10-3
NO2-=7 x 10-1

pH range:pH 3-8
(at 10-3 mol/l)



TOP
Copyright (C) 2008 HORIBA, Ltd. All rights reserved.

Copyright © 2008 HORIBA, Ltd. All rights reserved. The information shown on this document may be modified without notice. Refer to the original web page for update. The page was copied from:
 