EMGA-620W Series
Specifications

OXYGEN/NITROGEN ANALYZER
EMGA-620W Series

Specifications
Applications
O and/or N in ferrous and non-ferrous metals
and electronic materials
Oxygen/Nitrogen(EMGA-620W)
Nitrogen(EMGA-622W)
Oxygen(EMGA-623W)
Principle
Inert gas fusion in impulse furnace
O(Oxygen):NDIR(Non Dispersive Infra Red absorption)
N(Nitrogen):TCD(Thermal Conductivity Detection)
Range
O:0 - 0.1 wt%/0 - 1,000 wt ppm
N:0 - 0.5 wt%/0 - 5,000 wt ppm
* Specify wt% or wt ppm when ordering.
* The range may be expanded to 0 - 100 wt% by reducing the sample mass.
Sample required
1.0 g for regular analysis
Accuracy
(repeatability)
Accuracy
Sensitivity
(minimum reading)
0.000001 wt%/0.01 wt ppm or 0.0000001 wt%/0.001 wt ppm
Analysis time
Approx. 40 sec from gas extraction to measurement result display
* Depending on sample and extraction conditions.
Analysis conditions
5 combinations of the following conditions can be preprogrammed
Extraction conditions:
Furnace control:Electrical power control
circle Wattage setting
circleTemperature setting
*10 combinations of initial value, final value and time can be preprogrammed.
Operation mode:2 operation modes are available depending on the sample
circleBlock mode
circlePower mode
Integration conditions:
3 integration methods are available. Waiting time before integration can be set at 0 - 20,000 sec
circlePreset time integration
circle Integration up to reach comparator level
circlePreset time integration or integration up to reach comparator level, whichever is shorter
Calibration
Multi-point appoximation by least squares method
circle One-point or multi-point (up to 30 points) calibration using standard sample
circleOne-point or multi-point (up to 30 points) calibration using standard gas - (option, standard and for 621W)
circleEasy calibration using regular analysis data
Display
Colour TFT screen:
circle Interactive display for setting the conditions of analysis and maintenance
circleAnalysis data including results and conditions
circleAlarm messages
circleO/N extraction curve
circleTemperature curve
circleIntegration curve
Printer:colour printer
circleAnalysis data including results and conditions (automatically printed out after each analysis is completed.)
circlePreprogrammed analysis conditions
circleO/N extraction curve
circleTemperature curve
circleIntegration curve
Cooling method
Internal cooling water circulation method
Functions
circleData storage:
 
Pre-programmed sample data up to 100 items
Analysis results up to 1000 per file
circleFurnace control:10 furnace control patterns can be preprogrammed
circleFurnace temperature hold:The furnace temperature can be held at will when operating in interrupt analysis
circleInterrupt analysis
circleSetting and checking of analysis control range
circleSample fusion status monitoring
circleRS-232C output
circleSelf diagnosis:
 
Reagent needs replacing
Dust-filter needs changing
Cooling water condition fault
Furnace condition fault
Operating gas condition fault
Carrier gas condition fault
Others