Application
|
Oil contamination of fresh
and salt water; soil analysis
|
Principle
|
Solvent extraction, NDIR analysis
(Infrared spectroscopy)
|
Detector
|
Pyroelectric sensor
|
Measuring Range
|
0 to 200 mg/L
|
Resolution
|
0 to 99.9 mg/L; 0.1 mg/L
100 to 200 mg/L; 1 mg/L
|
Repeatability
|
0 to 9.9 mg/L; ± 0.2
mg/L ± 1 digit
10.0 to 99.9 mg/L; ± 2 mg/L ± 1 digit
100 to 200 mg/L; ± 4 mg/L ± 1 digit
|
Measurement
|
Automatic after introduction
of the oily water sample or manual operation, selectable
|
Calibration
|
Automatic zero and span calibration
after the oily water sample is introduced to the instrument.
|
Extraction Solvent
|
HORIBA H-997 solvent
|
Sample/Solvent Volume
|
20 mL/10 mL
|
Extraction Method
|
Built-in extractor for oily
water samples
|
Display
|
Measured value; 3 digits LCD
with backlit
Message; Character display LCD with backlit (16 x 2 characters)
|
Functions
|
Data memory up to 50 data,
Self diagnostics, Stable type display, Interactive operation, Calendar
clock
|
Output
|
RS-232C and parallel printer
port
|
Ambient Temperature
|
0 to 40 °C
|
Power Requirement
|
100 to 240 V AC ±10%,
50/60 Hz
100 to 120 V AC; approx. 80 VA
200 to 240 V AC; approx. 120 VA
|
Dimensions
|
200(W) x 315(D) x 342(H) mm
7.8(W) x 12.4(D) x 13.5(H) in
|
Mass
|
Approx. 7 kg/15.5 lb
|