Water Quality Monitoring System
U-20XD Series
| |
U-22XD SET
|
W-22XD
|
W-23XD
|
|
Control Unit
|
Water proof construction
|
IP-67
|
|
( 
)
option
|
( 
)
option
|
|
Multi-Probe *1
|
Measurement temperature
|
0~55°C
|
(to30m)
|
|
|
|
Storage temperature
|
-5~60°C
|
|
Measurement depth *2
|
to 100 m
|
|
Maximum probe size
|
95 mm
|
|
Probe length
|
430 mm
|
|
Continuous use *3
|
30days
|
|
Data logging
|
|
|
Manual data memory (2880
items)
|
|
|
Automatic calibration
|
|
|
pH
*Two-point calibration
*Automatic temperature compensation
|
Measurment Principle
|
Glass electrode method
|
|
( 
)
option
|
( 
)
option
|
|
Range
|
pH 0~14
|
|
Resolution
|
0.01 pH
|
|
Repeatability
|
±0.05 pH
|
|
Accuracy
|
±0.1 pH
|
|
Dissolved oxygen
*Salt correction
(0 to 40 ppt /automatic)
*Automatic temperature compensation
|
Measurment Principle
|
Diaphragm galvanic battery
method
|
|
( 
)
option
|
( 
)
option
|
|
Range
|
0~19.99 mg/L
|
|
Resolution
|
0.01 mg/L
|
|
Repeatability
|
±0.1 mg/L
|
|
Accuracy
|
±0.2 mg/L
|
|
Conductivity
*Auto range
*Automatic temperature
conversion (25°C)
*SI units
|
Measurment Principle
|
4 AC electrode method
|
|
|
|
|
Range
|
0~9.99 S/m
|
|
Resolution
|
0.1%F.S
|
|
Repeatability
|
±1%
|
|
Accuracy
|
±3%
|
|
Salinity
|
Measurment Principle
|
Conductivity conversion
|
|
|
|
|
Range
|
0~4%
|
|
Resolution
|
0.01%
|
|
Repeatability
|
±0.1%
|
|
Accuracy
|
±0.3%
|
|
Total Dissolved Solids (TDS)
*Conversion factor setting
|
Measurment Principle
|
Conductivity conversion
|
|
|
|
|
Range
|
0~100 g/L
|
|
Resolution
|
0.1%F.S
|
|
Repeatability
|
±2 g/L
|
|
Accuracy
|
±5 g/L
|
Seawater specific gravity
*Display  t,
 0,
 15
|
Measurment Principle
|
Conductivity conversion
|
|
|
|
|
Range
|
0~50  t
|
|
Resolution
|
0.1  t
|
|
Repeatability
|
±2  t
|
|
Accuracy
|
±5  t
|
|
Temperature
|
Measurment Principle
|
Thermistor method
|
|
|
|
|
Range
|
0~55°C
|
|
Resolution
|
0.01°C
|
|
Repeatability
|
±0.3°C
|
|
Accuracy
|
±1.0°C
|
|
Turbidity
*Unit selection
|
Measurment Principle
|
Penetration and scattering
method
|
|
|
|
|
Range
|
0~800 NTU
|
|
Resolution
|
0.1 NTU
|
|
Repeatability
|
±3%
|
|
Accuracy
|
±5%
|
|
Water depth
|
Measurment Principle
|
Pressure method
|
|
|
|
|
Range(NTU or mg/L)
|
0~100m
|
|
Resolution
|
0.1 m
|
|
Repeatability
|
±3%
|
|
Accuracy
|
±5%
|
|
Oxidation reduction potential(ORP)
|
Measurment Principle
|
Platinum electrode method
|
|
( 
)
option
|
( 
)
option
|
|
Range
|
±1999 mV
|
|
Resolution
|
1 mV
|
|
Repeatability
|
±5 mV
|
|
Accuracy
|
±15 mV
|
|
Ion
*Auto range
|
Measurment Principle
|
Ion electrode method
|
-
|
-
|
( 
)
option
( 
)
option
( 
)
option
|
|
Resolution
|
0.1%F.S
|
|
Repeatability
|
±5%
|
|
Accuracy
|
±10%
|
|
Nitric acid ion
|
Range
|
NO3-:
0.62~62,000 mg/L(pH 3~7)
|
|
Chloride ion
|
Cl-: 0.4~35,000
mg/L(pH 3~11)
|
|
Calcium ion
|
Ca2+:
0.4~40,080 mg/L(pH 5~11)
|
|
Fluoride ion
|
F-: 0.02~19,000
mg/L(pH4~10:20 mg/L)
|
|
Potassium ion
|
K+: 0.04~39,000
mg/L(pH5~11:3.9 mg/L)
|
|
Ammonia
|
NH3: 0.1~1,000
mg/L(pH 12 or more)
|
|
Simultaneously measurable parameters
|
|
|
10
|
10
|
13
|
|
Note: The accuracy rating value is obtained from
measurements at an intermediate point of the standard solution after
two-point calibration (at room temperature and pressure).
The repeatability and accuracy rating percentages are based on the
full scale (except for salinity).
*1: Organic solvents, strong acids, and strong alkaline solvents
cannot be measured.
*2: The maximum depth for ion measurements are 100 m for nitric
acid ion, chioride ion, fluoride ion, 15 m for calcium ion, ammonia,
and 3 m for potassium ion.
*3: Based on the data measured automatically at 15 minutes intervals.
The battery life taken into account.
Periodical maintenance and calibration is necessary when a lot of
shellfishes and seaweeds exist at the measurement point.
|
 |