Parameter
|
CRP
|
|
Sample species
|
whole blood, blood plasma, blood serum
|
|
Measurement principle
|
Latex immunoturbidmetry
|
|
Throughput
|
260 sec/sample (normal mode and wide
range mode)
|
|
Sample volume
|
8µL (normal mode)
4µL (wide range mode)
|
|
Linearity range
|
Whole blood: 0.2-10.0mg/dL (normal mode)
10.1-20.0mg/dL (wide range mode)
Blood plasma, serum: 0.2-7.0mg/dL (normal mode)
7.1-14.0mg/dL (wide range mode)
|
|
Displayed linearity range
|
0.0-10.0mg/dL (normal mode)
5.0-20.0mg/dL (wide range mode)
|
Repeatability |
<10% (1.0mg/dL) |
|
For measurement range
|
<4% (1.0-10.0mg/dL, normal mode)
<8% (10.1-20.0mg/dL, wide range mode)
|
|
Printer
|
built-in
|
|
LIS
|
output through RS-232C
|
|
Measurement data memory
|
last 100 sample
|
|
Dimensions
|
289*415*410mm(W*H*D)
|
|
Weight
|
approx. 18 kg
|