Measurement object
|
Ambient trace ammonia gas in clean rooms
|
Measurement principle
|
CRDS (Cavity Ring Down Spectroscopy)
|
Range
|
0 to 500 ppb
|
Zero noise
|
0.06 ppb or less (5 min. average mode over 24 hr. x σ)
|
Lower detection limit
|
0.1 ppb or less (30 min. average mode over 24 hr. x 3σ)
0.2 ppb or less (5 min. average mode over 24 hr. x 3σ)
|
Minimum interval
|
Approx. 5 sec.
|
Zero drift
|
±0.1 ppb/day (peak value of 50 min. average data)
|
Zero return time
|
60 min. or less
(Time to reach 0.2 ppb in 5 min. average mode after measuring sample gas of approx. 10 ppb)
|
Coupling
|
Outer diameter: 6.0 mm, shared-use: 6.35 mm,
fluorocarbon polymer
|
Outputs
|
Analog (concentration value)
Alarm (error, H level, HH level)
|
Stand
|
600 (W) x 835 (D) x 840 (H) mm (not incl. monitor, handles)
23.6 (W) x 32.9 (D) x 33.1 (H) in
|
Mass
|
110 kg, 242 lb
|
Power supply
|
100 V AC±10 V, 50/60 Hz
700 V A (max. consump.)
|
Sample gas conditions
|
Flow rate: 0.4 to 10 L/min
Temperature: 15 to 35°C
|