Model
|
CS-151F1
|
CS-152F1
|
CS-150F1
|
 |
 |
 |
 |
|
Target of
measurement
|
SC-1
(ammonia/hydrogen peroxide aqueous solution)
|
SC-2
(hydrochloric acid/hydrogen peroxide aqueous solution)
|
SPM
(sulfuric acid/hydrogen peroxide aqueous solution)
|
|
Multi-monitoring with one unit of up to 4 types of chemical solutions or ranges possible.
|
|
Measurement principle
|
Absorption spectrometry
|
|
Concentration computation principle
|
Temperature-compensated multivariate analysis
|
|
Measurement range
(mass %)
|
NH3 : 0.00% to 1.00%
H2O2 : 0.00% to 5.00%
H2O : 94.0% to 100.0%
|
HCI : 0.00% to 2.00%
H2O2 : 0.00% to 2.00%
H2O : 96.0% to 100.0%
|
H2SO4 : 70.0% to 96.0%
H2O2 : 0.00% to 10.00%
H2O : 4.0% to 30.0%
|
|
Repeatability (mass %)
(within ±1% room temperature)
|
NH3 : ±0.15%
H2O2 : ±0.30%
H2O : ±1.5%
|
HCI : ±0.15%
H2O2 : ±0.15 %
H2O : ±1.5%
|
H2SO4 : ±0.50%
H2O2 : ±0.50%
H2O : ±1.5%
|
|
Sample solution temperature
|
20 to 80 : (when using Inline cell) 20 to 30 : (when using cell unit)
|
|
Measurement cycle
|
Approx. 3 seconds
|
|
Content of
LCD and input
|
|
1st component: measured concentration value (%)
2nd component: measured concentration value (%)
3nd component: measured concentration value (%)
1st and 2nd components: concentration alarm status
|
Error code
Parallel concentration alarm output invalid status
Serial/parallel status
Background compensation data
|
|
|
RS-232C serial input
|
|
Serial/parallel status switching
Measurement data transfer request
1st and 2nd component concentration alarm setting
|
Parallel concentration alarm output status
(valid/invalid)
Target measurement solution switching
|
|
|
Parallel input
|
|
12 to 30 V DC (photocoupler insulated)
|
Parallel concentration alarm output
Target measurement solution switching
Background compensation
|
|
|
Parallel output
|
|
Open collector output (photocoupler insulated)
Max. current when on:
DC 5 mA (no internal protection resistor)
Max. ignition voltage when off: 30 V DC
|
1st component concentration alarm, 4 types
2nd component concentration alarm, 4 types
Device error
Measuring
Target measurement solution type
Background compensation status
|
|
|
Solution temperature input
|
Terminal block connection (M3 terminal screws)
Platinum resistance temperature sensor: Pt100 (Pt100 sensor should be prepared by customers)
|
|
Analog output
|
1st component concentration alarm, 4 types: 4 to 20 mA
2nd component concentration alarm, 4 types: 4 to 20 mA
* We also support other measurement ranges. Please specify when ordering.
|
|
Fiber optic cable
|
Dedicated CS-100F1 Series fiber optic cable
(25 ±15: installation temperature)
|
|
ample cell
|
We can supply dedicated sample cells (inline cell or cell unit) for various kinds of chemical solutions.
|
|
Power supply
|
100 V to 230 V AC (single phase) ±10%, 50/60 Hz, approx. 100 VA
|
|
Dimensions
|
205 (W) x 329 (D) x 269 (H) mm 8.1 (W) x 13.0 (D) x 10.6 (H) in (excluding protruding part)
|
|
Weight
|
Approx. 11 kg
|