Japan Worldwide
News & EventsInvestorsProcurementCompany HomeContactsSite MapJapanese
Products & Support Automotive Process & Environmental Medical-Diagnostics Semiconductor Scientific Applications
Measurement Types
Product List
 HOME >> Products & Support
Semiconductor Instruments & SystemsProducts
product image
CS-100F1
Specifications
Main Features
External dimensions
Related Products

Fiber Optic Type
CHEMICAL SOLUTION
CONCENTRATION MONITOR
CS-100F1 Series

Specifications
Model
CS-151F1
CS-152F1
CS-150F1
Target of
measurement
SC-1
(ammonia/hydrogen peroxide aqueous solution)
SC-2
(hydrochloric acid/hydrogen peroxide aqueous solution)
SPM
(sulfuric acid/hydrogen peroxide aqueous solution)
Multi-monitoring with one unit of up to 4 types of chemical solutions or ranges possible.
Measurement principle
Absorption spectrometry
Concentration computation principle
Temperature-compensated multivariate analysis
Measurement range
(mass %)
NH3 : 0.00% to 1.00%
H2O2 : 0.00% to 5.00%
H2O : 94.0% to 100.0%
HCI : 0.00% to 2.00%
H2O2 : 0.00% to 2.00%
H2O : 96.0% to 100.0%
H2SO4 : 70.0% to 96.0%
H2O2 : 0.00% to 10.00%
H2O : 4.0% to 30.0%
Repeatability (mass %)
(within ±1% room temperature)
NH3 : ±0.15%
H2O2 : ±0.30%
H2O : ±1.5%
HCI : ±0.15%
H2O2 : ±0.15 %
H2O : ±1.5%
H2SO4 : ±0.50%
H2O2 : ±0.50%
H2O : ±1.5%
Sample solution temperature
20 to 80 : (when using Inline cell)  20 to 30 : (when using cell unit)
Measurement cycle
Approx. 3 seconds
Content of
LCD and input
• 1st component: measured concentration value (%)
• 2nd component: measured concentration value (%)
• 3nd component: measured concentration value (%)
• 1st and 2nd components: concentration alarm status
• Error code
• Parallel concentration alarm output invalid status
• Serial/parallel status
• Background compensation data
RS-232C serial input
• Serial/parallel status switching
• Measurement data transfer request
• 1st and 2nd component concentration alarm setting
• Parallel concentration alarm output status
(valid/invalid)
• Target measurement solution switching
Parallel input
12 to 30 V DC (photocoupler insulated)
• Parallel concentration alarm output
• Target measurement solution switching
• Background compensation
Parallel output
Open collector output (photocoupler insulated)
Max. current when on:
DC 5 mA (no internal protection resistor)
Max. ignition voltage when off: 30 V DC
• 1st component concentration alarm, 4 types
• 2nd component concentration alarm, 4 types
• Device error
• Measuring
• Target measurement solution type
• Background compensation status
Solution temperature input
Terminal block connection (M3 terminal screws)
• Platinum resistance temperature sensor: Pt100 (Pt100 sensor should be prepared by customers)
Analog output
• 1st component concentration alarm, 4 types: 4 to 20 mA
• 2nd component concentration alarm, 4 types: 4 to 20 mA
* We also support other measurement ranges. Please specify when ordering.
Fiber optic cable
Dedicated CS-100F1 Series fiber optic cable
(25 ±15: installation temperature)
ample cell
We can supply dedicated sample cells (inline cell or cell unit) for various kinds of chemical solutions.
Power supply
100 V to 230 V AC (single phase) ±10%, 50/60 Hz, approx. 100 VA
Dimensions
205 (W) x 329 (D) x 269 (H) mm 8.1 (W) x 13.0 (D) x 10.6 (H) in (excluding protruding part)
Weight
Approx. 11 kg
Note : Please consult us regarding concentration monitors for chemical solutions other than those given above.

TOP
Copyright (C) 2008 HORIBA, Ltd. All rights reserved.

Copyright © 2008 HORIBA, Ltd. All rights reserved. The information shown on this document may be modified without notice. Refer to the original web page for update. The page was copied from:
 
Explore the future HORIBA