CS-131
Specifications

SC-1 monitor

CS-131

Specifications

Model specifications CS-131: No cooling unit
CS-131C: Integrated cooling unit
Measurement sample SC-1 (ammonia/hydrogen peroxide aqueous solution)
Measurement principle Absorption spectrometry
Concentration computation principle Temperature-compensated multivariate analysis
Measurement range NH3: 0.00 to 5.00%
H2O2: 0.0 to 10.00%
H2O: 85.0 to 100.0%
Please consult HORIBA for information concerning other measurement ranges.
Reproducibility (Peak to peak) NH3: ± 0.25%
H2O2: ± 0.3%
H2O: ± 1.5%
Chemical conditions CS-131:
Temperature: 20 to 30°C
Flow rate: 20 to 60 mL/min.
Pressure: 0.10 to 0.20 MPa
CS-131C:
Temperature: 20 to 80°C
Flow rate: 20 to 40 mL/min.
Pressure: 0.10 to 0.20 MPa
Measurement cycle Approx. 2 seconds minimum
(not including chemical replacement and cooling times)
Equipment temperature 20 to 25
Operation panel 1) Temperature level display: Ammonia, hydrogen peroxide, water
2) Alarm display: Ammonia, hydrogen peroxide: Concentration low-low limit, concentration low limit, concentration high limit, concentration high-high limit
3) Alarm value settings: Ammonia, hydrogen peroxide: Concentration low-low limit, concentration low limit, concentration high limit, concentration high-high limit
Serial input/output RS-232C
1) Concentration data: Ammonia, hydrogen peroxide
2) Alarm value settings: Ammonia, hydrogen peroxide: Concentration low-low limit, concentration low limit, concentration high limit, concentration high-high limit Parallel input 12 to 30 V DC (photocoupler insulated) Parallel concentration alarm (ON/OFF)
Parallel output Maximum current when open collector output is ON: 5 mA
Maximum voltage when open collector output is OFF: 30 V
1) Alarm output: Ammonia, hydrogen peroxide: Concentration low-low limit, concentration low limit, concentration high limit, concentration high-high limit
2) Device error
3) Liquid leakage
Analog output 4 to 20 mA (2 outputs: NH3 concentration, H2O2 concentration) *Regardless of the range specifications, the output of 4 ~ 20 mA can be made to correspond to 0 ~ 10% concentrations.
(Please specify when ordering.)
Power supply 24 V DC ± 10%, 2 A
Dimensions CS-131: 205 (W) x 329 (D) x 269 (H) mm
CS-131C: 291 (W) x 329 (D) x 269 (H) mm
Weight CS-131: Approx. 11 kg
CS-131C: Approx. 12 kg
Connection pipe size For input and exhaust: Outer diameter 3 mm (inner diameter 2 mm)
Pipe connection Nippon Pillar Super type
Please consult HORIBA concerning any specifications not listed above.