CS-150
Specifications
Sample configuration
External dimensions

SPM Monitor
CS-150

Specifications
Model
CS-150
CS-150C
Model specifications
No cooling unit
Integrated cooling unit
Measurement sample
SPM (Sulfuric acid/Hydrogen peroxide aqueous solution)
Measurement principle
Absorption spectrometry
Concentration
computation principle
Temperature-compensated multivariate analysis
Measurement range
H2SO4: 50.0 to 90.0 %
H2O2: 0.00 to 10.00 %
H2O: 0.0 to 50.0 %
Reproducibility
(Peak to peak)
H2SO4: ±0.5 %
H2O2: ±0.50 %
H2O: ±1.5 %
Chemical
conditions
Temperature
20 to 30 °C
20 to 60 °C
Flow rate
20 to 60 ml/min.
20 to 40 ml/min.
Pressure
0.10 to 0.20 MPa
Measurement cycle
Approx. 3 seconds minimum
Equipment temperature
20 to 25 °C
Operation panel
1)
Temperature level display:
Sulfuric acid, Hydrogen peroxide, Water
2)
Alarm display (Sulfuric acid, Hydrogen peroxide):
Concentration low-low limit, Concentration low limit,
Concentration high limit, Concentration high-high limit
3)
Alarm value settings (Sulfuric acid, Hydrogen peroxide):
Concentration low-low limit, Concentration low limit,
Concentration high limit, Concentration high-high limit
Serial input/output
RS-232C
1)
Concentration data:
Sulfuric acid, Hydrogen peroxide, Water
2)
Alarm value settings (Sulfuric acid, Hydrogen peroxide):
Concentration low-low limit, Concentration low limit,
Concentration high limit, Concentration high-high limit
Parallel input
12 to 30 V DC (photocoupler insulated)
Parallel concentration alarm (ON/OFF)
Parallel output
Maximum current when open collector output is ON:
5 mA
Maximum voltage when open collector output is OFF:
30 V
1)
Alarm output: Sulfuric acid, Hydrogen peroxide:
Concentration low-low limit, Concentration low limit,
Concentration high limit, Concentration high-high limit
2)
Device error (including liquid leakage)
3)
Under measurement
Analog output
4 to 20 mA
(2 outputs: H2SO4: 50.0 to 100.0 %/H2O: 0.00 to 10.00 %)
The analog output can be prepared to correspond to each
requirement by users. (Please specify when ordering.)
Power
24 V DC ±10%, 2 A
Dimensions
205 (W) x 329 (D) x 269 (H) mm
8.1 (W) x 13 (D) x 10.6 (H) in
291 (W) x 329 (D) x 269 (H) mm
11.5 (W) x 13 (D) x 10.6 (H) in
Mass
Approx. 11 kg
Approx. 12 kg
Connection pipe size
For inlet and outlet: Outer diameter 6 mm (inner diameter 4 mm)
Pipe connection
Nippon Pillar Super type
Please consult HORIBA concerning any specifications not listed above.