CS-152
Specifications
Sample configuration
External dimensions

SC-2 Monitor
CS-152

Specifications
Model
CS-152
CS-152C
Model specifications
No cooling unit
Integrated cooling unit
Measurement sample
SC-2 (Hydrochloric acid/Hydrogen peroxide aqueous solution)
Measurement principle
Absorption spectrometry
Concentration
computation principle
Temperature-compensated multivariate analysis
Measurement range
HCl: 0.00 to 2.00 %/0.00 to 10.00 %
H2O2: 0.00 to 2.00 %/0.00 to 10.00 %
H2O: 96.0 to 100.0 %/80.0 to 100.0 %
Reproducibility
(Peak to peak)
HCl: ±0.15 %/±0.20 %
H2O2: ±0.15 %/±0.20 %
H2O: ±1.5 %/±1.5 %
Chemical
conditions
Temperature
20 to 30 °C
20 to 80 °C
Flow rate
20 to 60 ml/min.
20 to 40 ml/min.
Pressure
0.10 to 0.20 MPa
Measurement cycle
Approx. 3 seconds minimum
Equipment temperature
20 to 25 °C
Operation panel
1)
Temperature level display:
Hydrochloric acid, Hydrogen peroxide, Water
2)
Alarm display (Hydrochloric acid, Hydrogen peroxide):
Concentration low-low limit, Concentration low limit,
Concentration high limit, Concentration high-high limit
3)
Alarm value settings (Hydrochloric acid, Hydrogen peroxide):
Concentration low-low limit, Concentration low limit,
Concentration high limit, Concentration high-high limit
Serial input/output
RS-232C
1)
Concentration data:
Hydrochloric acid, Hydrogen peroxide, Water
2)
Alarm value settings (Hydrochloric acid, Hydrogen peroxide):
Concentration low-low limit, Concentration low limit,
Concentration high limit, Concentration high-high limit
Parallel input
12 to 30 V DC (photocoupler insulated)
Parallel concentration alarm (ON/OFF)
Parallel output
Maximum current when open collector output is ON:
5 mA
Maximum voltage when open collector output is OFF:
30 V
1)
Alarm output (Hydrochloric acid, Hydrogen peroxide):
Concentration low-low limit, Concentration low limit,
Concentration high limit, Concentration high-high limit
2)
Device error (including liquid leakage)
3)
Under measurement
Analog output
4 to 20 mA (2 outputs: HCl concentration, H2O concentration
The analog output can be prepared to correspond to each
requirement by users. (Please specify when ordering.)
Power
24 V DC ±10%, 2 A
Dimensions
205 (W) x 329 (D) x 269 (H) mm
8.5 (W) x 13 (D) x 10.6 (H) in
291 (W) x 329 (D) x 269 (H) mm
11.5 (W) x 13 (D) x 10.6 (H) in
Mass
Approx. 11 kg
Approx. 12 kg
Connection pipe size
For inlet and outlet: Outer diameter 3 mm (inner diameter 2 mm)
Pipe connection
Nippon Pillar Super type
Utilities
Instrumentation air for driving air valve:
0.3 to 0.5 MPa (Piping outer diameter: 4 mm)
Please consult HORIBA concerning any specifications not listed above.