Model name |
FG-110A |
FG-120A |
Measurement principle |
FTIR method |
Detector |
Liquid-nitrogen-cooled MCT detector |
Electronically-cooled MCT detector |
Measurement wavenumber range |
Standard: 5000 to 700 cm-1
Option: 5000 to 600 cm-1
|
5000 to 900 cm-1 |
Cell optical path length
(cell volume) |
Single cell:
0.01 m (6.2 mL), 0.1 m (62 mL), 0.8 m (220 mL),
2.4 m (220 mL), 10 m (1800 mL)
Dual cell:
0.01 m + 2.4 m, 0.01 m + 0.8 m, 0.1 m + 2.4 m |
Cell window material |
Standard: BaF2, Option: ZnSe |
O-ring material |
Standard: Viton, Option: Kalrez |
Piping |
Standard: 1/4 inch Swagelok, Option: 1/4 inch VCR |
Resolution |
0.5, 1, 2, 4, 8, 16 (cm-1) |
Measurement cycle |
Approx. 0.6 seconds (Varies depending on measurement conditions) |
Ambient temperature |
10 to 30°C |
Humidity |
70% Rh max. |
Power |
100 to 240 V ± 10 V AC, 50/60 Hz |
Power consumption |
240 VA (2.4 m cell) |
250 VA (2.4 m cell) |
Nitrogen gas |
For analyzer purging: Approximately 5 L/min, continuous
For calibration: Approximately 5 L/min, as needed
Nitrogen gas conditions: Purity 99.99% or higher
Degree of saturation -30°C or lower
No dust or mist
|
Liquid nitrogen |
Approx, 500 to 600 mL
(maintained for 8 to 10 hours) |
-- |
Mass |
Single cell type: 40 kg (2.4 m cell), Dual cell type: 50 kg |
Applicable regulation |
CE marking, FCC Part 15 (Class A) |