Japan Worldwide
News & EventsInvestorsProcurementCompany HomeContactsSite MapJapanese
Products & Support Automotive Process & Environmental Medical-Diagnostics Semiconductor Scientific Applications
Measurement Types
Product List
 HOME >> Products >> Semiconductor Instruments & Systems >> FG-100A Series
Semiconductor Instruments & SystemsProducts
product image
FG-100A Series
Specifications
Control Unit
Sampling Unit
Suitable for diverse applications
Flow Schematics
Typical measurement results
Dimensional Outlines
World-wide network ensures full after-sales service
Related product

FTIR Gas Analyzer

FG-100A Series

Specifications
Model name
FG-110A
FG-120A
Measurement principle
FTIR method
Detector
Liquid-nitrogen-cooled MCT detector
Electronically-cooled MCT detector
Measurement wavenumber range
Standard: 5000 to 700 cm-1
Option: 5000 to 600 cm-1
5000 to 900 cm-1
Cell optical path length
(cell volume)
Single cell:
0.01 m (6.2 mL), 0.1 m (62 mL), 0.8 m (220 mL),
2.4 m (220 mL), 10 m (1800 mL)
Dual cell:
0.01 m + 2.4 m, 0.01 m + 0.8 m, 0.1 m + 2.4 m
Cell window material
Standard: BaF2, Option: ZnSe
O-ring material
Standard: Viton, Option: Kalrez
Piping
Standard: 1/4 inch Swagelok, Option: 1/4 inch VCR
Resolution
0.5, 1, 2, 4, 8, 16 (cm-1)
Measurement cycle
Approx. 0.6 seconds (Varies depending on measurement conditions)
Ambient temperature
10 to 30°C
Humidity
70% Rh max.
Power
100 to 240 V ± 10 V AC, 50/60 Hz
Power consumption
240 VA (2.4 m cell)
250 VA (2.4 m cell)
Nitrogen gas
For analyzer purging: Approximately 5 L/min, continuous
For calibration: Approximately 5 L/min, as needed
Nitrogen gas conditions: Purity 99.99% or higher
Degree of saturation -30°C or lower
No dust or mist
Liquid nitrogen
Approx, 500 to 600 mL
(maintained for 8 to 10 hours)
--
Mass
Single cell type: 40 kg (2.4 m cell), Dual cell type: 50 kg
Applicable regulation
CE marking, FCC Part 15 (Class A)


TOP
Copyright (C) 2008 HORIBA, Ltd. All rights reserved.

Copyright © 2008 HORIBA, Ltd. All rights reserved. The information shown on this document may be modified without notice. Refer to the original web page for update. The page was copied from:
 
Explore the future HORIBA